eCommons

 

Shear Stress Induced Thrombogenicity of a Trileaflet Mechanical Heart Valve

dc.contributor.authorGregan, Chris
dc.contributor.authorMoy, David
dc.contributor.authorNewberger, Nicole
dc.contributor.authorSiomos, Matthew
dc.date.accessioned2019-09-11T17:45:49Z
dc.date.available2019-09-11T17:45:49Z
dc.date.issued2019-05
dc.description.abstractTrileaflet mechanical valves are a popular topic in industry R&D due to their potential improvements to hemodynamic performance relative to industry-standard bileaflet mechanical valves and their mechanical durability relative to tissue valves. Novostia is currently attempting to bring their Lapeyre-Triflo trileaflet valve through FDA approvals, demonstrating a clear need and viability for the design concept. While some analysis of this new design has been done, there is a lack of research into the effect of the new leaflet geometry on peak shear stresses in the flow, which impact the thrombogenicity of the valve. Thrombosis is one of the leading causes of complications associated with mechanical heart valves on the market today. While many different factors contribute to the thrombogenicity of a heart valve, high shear stresses in the flow are classically considered to be a significant contributing factor due to the platelet damage that occurs in high shear stress regions. Specifically, our team examined whether shear stresses and resultant platelet damage are increased relative to classic bileaflet valve designs when the valve is in the open position at peak systole. The triangular leaflet geometry contains a sharp trailing edge, which could increase the shear stresses, and the design also introduces an additional region of flow (four flow regions rather than three), which brings potential for further impacts to shear stress downstream of the leaflets.en_US
dc.identifier.urihttps://hdl.handle.net/1813/66939
dc.language.isoen_USen_US
dc.subjectThrombogenicityen_US
dc.subjectTrileafleten_US
dc.subjectHeart Valveen_US
dc.titleShear Stress Induced Thrombogenicity of a Trileaflet Mechanical Heart Valveen_US
dc.typepresentationen_US

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Group 3 - Shear Stress Induced Thrombogenicity of a Trileaflet Mechanical Heart Valve.pdf
Size:
2.53 MB
Format:
Adobe Portable Document Format
Description: